LF1030 HYDRAULIC PATIENT LIFT / LF1040 ELECTRIC PATIENT LIFT OWNER'S MANUAL, 10/03

13

Note: Steps 10-13 apply to LF1040 Electric Patient

Lift only.

10. Attach control box using 5mm cap screws as shown

at right.

11. Attach battery charger using 5mm cap screws as

shown at right.

12. Clip battery onto unit.

13. See following Electrical connections section.

Electrical connections (LF1040 only)

The control box is the heart of the electrical system; into it plug the hand control, battery, and battery
charger (depending on battery charger options). Ensure that all plugs are seated firmly in the proper
receptacles.

When standing behind the patient lift, the hand control connection is on the far left of the control box.
Ensure that the marks on the control box and the hand control plug are aligned, and then firmly insert
the hand control plug. Immediately to the right of the hand control connection is the motor connection.
Insert the hand control plug firmly in its receptacle (this plug has nothing to align).

Charging the battery (LF1040 only)

The battery used on the LF1040 has an audible alarm that sounds when the battery charge has been
depleted by 50%. Charge the battery as soon as possible upon hearing this alarm.

Note: This patient lift uses a lead acid battery. Charge the battery fully before initial use. To increase

the longevity of the battery, keep the battery fully charged when not in use. Battery charging time
is approximately four hours.

Warning: After the alarm sounds, there could still be sufficient battery power left to lift a
patient. Be aware that a lifted patient cannot be lowered when the battery is discharged.
Refer to the

Emergency Procedures section that follows for discharged battery operation.

Control box

Cap screw

Battery charger

Cap screw