4. Connect other end of air hoses to the pump.
5. Connect air hoses with fittings to the air chamber

openings on the pad (FIGURE 2).

6. Install sheet over pad and bed mattress.
7. Refer to USING/ADJUSTING THE PUMP in this

instruction sheet.

DO NOT, under any circumstances, cut or remove
the round grounding prong from any plug used
with or for Invacare products. Some devices are
equipped with three-prong (grounding) plugs for
protection against possible shock hazards. Where
a two-prong wall receptacle is encountered, it is
the personal responsibility and obligation of the
customer to contact a qualified electrician and
have the two-prong receptacle replaced with
a properly grounded three-prong wall recep-
tacle in accordance with the National Electrical
Code. If you must use an extension cord, use
ONLY a three-wire extension cord having the
same or higher electrical rating as the device
being connected.

WARNING

Using

1. Plug power cord into outlet.
2. Turn power switch to the ON position.
NOTE: The green light will be illuminated when the

pump is in the ON position.

USING/ADJUSTING THE PUMP
(FIGURE 3)

CARE AND MAINTENANCE

1. The pad can be cleaned by wiping it with a

germicidal disinfectant solution.

Pump

Knob

Power

Switch

FIGURE 3- USING/ADJUSTING THE PUMP

Air Chamber Openings

Hose

Pad

FIGURE 2- INSTALLING THE CAREGUARD APP-

ATTACHING HOSE TO PAD

3. Turn knob on pump to the maximum setting (clock-

wise).

NOTE: It will take approximately 15 minutes for the

pad to be fully inflated at the maximum p.s.i.
NOTE: Keep hose attached to the pad and the pump,

otherwise, deflation will occur.
4. Refer to ADJUSTING THE PUMP in this instruc-

tion sheet.

Adjusting

NOTE: The pressure of the pump can be adjusted

to provide maximum effectiveness and pressure

relief for the patient.
NOTE: The pressure can be adjusted with a range

from 1 to 8, with 1 being the lowest pressure settting.
NOTE: The pump will cycle every five(5) minutes

inflating and deflating the two air chambers.
1. Put patient on the pad.
2. Set the pressure to desired setting.
3. Slide the flat portion of your hand between the

pad and the matress of the bed.

NOTE: There should be a cushion of air between the pad

and the mattress of the bed. If not, check pump and pad

for leaks, tears, kinks, or breaks. If damage is found, DO

NOT use. Contact Carrier/Dealer for further instructions.
4. If pressure is unacceptable for patient, repeat

STEP 2 until desired pressure is reached.