Section 6- troubleshooting, Problem, Solution – Invacare IRC 1720 User Manual

Page 35: Turn the unit off for a minimum of 40, Minutes to allow the nebulizer to cool, The nebulizer does not operate, Check all power sources to the nebulizer, Push the on/off button again, Medication does not nebulize properly, Check for proper connection of the tubing adapter

Advertising
background image

SECTION 6—TROUBLESHOOTING

Part No 1130202

35

Stratos™Portable

SECTION 6—TROUBLESHOOTING

PROBLEM

SOLUTION

Nebulizer overheats. The

nebulizer has an internal

thermal protector to pre-

vent the device from over-

heating. If the internal

temperature reaches the

preset level, the unit will

automatically shut down, and

the power LED will turn

RED.

Turn the unit OFF for a minimum of 40

minutes to allow the nebulizer to cool.

The nebulizer does not

operate.

Check all power sources to the nebulizer.

Push the ON/OFF button again.

Medication does not nebu-

lize properly.

Check for proper connection of the tubing

adapter.

Make sure the nebulizer is in an upright position

and not tilted beyond 30º.

Make sure the nebulizer is not empty.

Clean the nebulizer as described in the manual.

Make sure the cone has been properly placed in

the nebulizer cup.

Replace the nebulizer.

Check the intake filter and replace it if it is dirty.

Nebulizer delivers no mist.

Make sure the cone is in the cup and properly

seated.