SECTION 5—OPERATING INSTRUCTIONS

Part No 1118353

21

Platinum

Series

Front Panel Indicators - Units with SensO

2

NOTE: For this procedure, refer to FIGURE 5.7.

The SensO

2

 feature monitors the purity of oxygen generated by 

the oxygen concentrator. If purity falls below factory preset 
standards, indicator lights on the control panel will illuminate.

FIGURE 5.7 Front Panel Indicators - Units with SensO2

Initial Startup of the Concentrator

NOTE: Concentrator may be used during the initial start warm‐up 
time (approximately 30 min.) while waiting for the O

2

 purity to 

reach maximum.

LABEL

SYMBOL

O

2

PURITY

INDICATOR LIGHTS

(LED)

SYSTEM OKAY
O

2

over 85% (± 2%)

GREEN Indicator Light

O

2

Between 73% (± 3%)

to 85% (±2%)

YELLOW Indicator light

A. YELLOW Solid

B. YELLOW Flashing Sensor Failure

Call a qualified technician. Have a

back-up supply of oxygen ready.

SYSTEM FAILURE
O

2

Below 73% (±3%)

RED Indicator Light
Continuous Audible Alarm
Sieve-GARD Compressor Shut-
down. Call a qualified technician.
Switch to a back-up supply of oxy-
gen immediately.

Call Supplier

Normal

Below Normal

OXYGEN

PURITY

RED

YELLOW

GREEN

O

2

Auto Shut - Down

Use Backup

O

2

Advertising