Compressor:

If the unit shuts down frequently, you may have a low voltage situation.

Low voltage can also be suspected when:

1. The motor does not get up to full power or speed.
2. Home fuses or circuit breakers activate when starting compressor.
3. Lights dim or remain dim when compressor is started.
4. Other motor operated appliances fail to operate properly or too

many motor operated appliances are on same circuit.

The motor of this compressor has a thermal overload protector. If the

motor should overheat, the overload protector will shut off the motor.

If the unit shuts down and cannot be restarted, the motor must be

replaced. Call your dealer immediately.

Extension Cords:

Make certain extension cord is in good condition. Make certain extension

cord wire is NOT LESS than 18 gauge for 25 feet, 16 gauge for 50 feet and

14 gauge for 100 feet.

MAINTENANCE

Only authorized or factory-trained personnel should perform preventative

maintenance.