Operation – Infinity M30 User Manual

Page 17

Advertising
background image

To start immediately, press START / STOP.

Follow the same programming procedures if a three or four stage program is desired.

Press NIGHT TIME (DELAY START) to start an overnight burnout. (See SAMPLE ON STAGE PROGRAM to pro-
gram NIGHT TIME (DELAY START) time.

NOTE:

When the temperature is set to a lower temperature than in the previous stage,
the furnace ignores the programmed HEAT RATE (except for "NO") and
automatically cools to the pre-set temperature. The time required for the cool
ing stage is determined by the pre-set temperature. The lower the temperature,
the more time needed to cool down.

OPERATION

HEAT RATE

TEMP

HOLD TIME

STAGE 1

STAGE 2

STAGE 3

STAGE 4

10°F (6°C)

1600°F (871°C)

1:00

NO

NO

*

*

*

*

* Though any number may appear, the unit is deactivated when “NO” is selected for the HEAT RATE.

20°F (11°C)

600°F (316°C)

30

16

SAMPLE THREE STAGE PROGRAM
(Program with Cooling)