SECTION 3—CLEANING AND DISINFECTING

Invacare

®

Select

22

Part No 1148073

2. Disassemble nebulizer chamber by turning 

counterclockwise and separating.

3. Fill two plastic containers or bowls:

A. Washing: Hot water/detergent solution.

B. Soaking: Hot water/vinegar solution (one part vinegar 

to three parts water).

4. Thoroughly clean the nebulizer and mouthpiece in hot 

water/detergent solution.

5. Remove from the water/detergent solution and rinse with 

clear hot tap water.

6. Soak in hot water/vinegar solution for 30 minutes.

7. Rinse with hot tap water again and air dry thoroughly.

NOTE: If using medical disinfectant cleaners, follow 
manufacturer’s instructions carefully.

8. Keep the outer surface of the tubing dust‐free by wiping 

regularly.

NOTE: The nebulizer air tubing does not have to be washed 
internally because only filtered air passes through the tubing.