Concentrator troubleshooting guide – Invacare PLATINUM 5 User Manual

Page 20

Advertising
background image

20

CONCENTRATOR

TROUBLESHOOTING GUIDE

MAINTENANCE/TROUBLESHOOTING

B.

CABINET

WARNING

UNPLUG the concentrator when cleaning. To avoid electrical

shock, DO NOT remove cabinet.

1. Clean the cabinet with a mild household cleaner and non-abrasive

cloth or sponge.

C.

HUMIDIFIER

NOTE: To clean the oxygen humidifier, follow the instructions provided by

the manufacturer. If none are provided, follow these steps:
1. Clean the humidifier every day.
2. Wash it in soapy water and rinse it with a solution of ten (10)

parts water and one (1) part vinegar.

3. Rinse thoroughly with hot water and refill with distilled water to

the level shown on the humidifier.

WARNING

DO NOT overfill Humidifier Bottle.

SOLUTION:

1. Insert plug into outlet.
2. Inspect house circuit

breakers or fuses. If
problem recurs, use a
different outlet.

3. Call Service Provider.

1a. Remove and clean

cabinet filters.

1b. Move oxygen concentrator at

least 7.5 cm (3 in) away from
walls, draperies or furniture.

2. Do not use extension

cords. Move to another
electrical outlet or circuit.

3. Call Service Provider.

PROBABLE CAUSE:

Main Power Loss:
1. Power cord NOT

plugged In.

2. No power at outlet.

3. Tripped circuit breaker.

System Failure:
1. Unit overheating due

to blocked air intake.

2. Insufficient power at

outlet.

3. Internal repairs required.

SYMPTOM:

Alarm:
SHORT BEEPS,
LONG PAUSE
Concentrator

NOT operating,

power switch ON.

Beep......Beep......
Alarm:
CONTINUOUS

Concentrator
NOT operating,
Power Switch ON.

Beeeeeeeep....