Lifescan Ultra 2 User Manual

Page 50

Advertising
background image

44

Customer Service

1 800 227-8862

44

Customer Service

1 800 227-8862

3. Match the code on the meter with the code on
the test strip vial

If the code on the meter does not match the
code on the test strip vial, press

or

to

match the code number on the test strip vial.
The new code number will flash on the display
for three seconds, then briefly stop flashing,
after which the display will advance to the

APPLY

BLOOD

screen.

If the codes already match, press

to go to the

APPLY BLOOD

screen. When you do not make a

change after five seconds, the display will
advance to the

APPLY BLOOD

screen.

4. Mark the test as a control solution test
Press

to change

APPLY BLOOD

to

APPLY

CONTROL

. You must mark the test before you

apply control solution. Once you have completed
the test, you cannot change the marking.

The meter is now ready to perform a control
solution test.

C

ontr

o

l

s

o

lution t

e

stin

g

DCO_06344501B_OU2_OB_USe.qxd 12/21/05 8:03 AM Page 44

Advertising