Unexpected messages and observations – Lifescan SureStep Pro User Manual

Page 97

Advertising
background image

85

If you cannot resolve a problem, call LifeScan Healthcare Professional Line:

1 800 524-7226 (USA)

1 888 353-0800 (Canada)

Unexpected Messages and Observations

MESSAGE

POSSIBLE

CAUSE

SOLUTION

Apply Blood Bedside unit is

ready for blood
application.

Obtain a drop of capillary
blood and apply it to center
of pink test square.
• If white pad becomes satu-

rated, you have applied
too much blood. Repeat
application with new test
strip.

• If white patches or streaks

appear on confirmation
dot, you have not applied
enough blood. Repeat
application with new test
strip.

Bedside Unit
Disabled

Bedside unit
deleted from
system or
unassigned.

Contact system
administrator for current
status of bedside unit.

Bedside Unit
Needs
Configuration

Initial data transfer
is required. Or,
configuration is
corrupted.

Transfer data.

BU.book Page 85 Monday, October 28, 2002 11:14 PM