Connect power, Set switch interface type – Polycom SpectraLink SCM408 User Manual

Page 35

Advertising
background image

Install SpectraLink 6000 System

PN: 1725-36097-001_L.doc

35

Connect power

1. Once the units have been properly grounded, connect the power

plug from the AC adapter to the jack labeled

PWR

on the MCU.

Use only the provided Class II AC Adapter with output 24V DC, 1A.

For installations with more than one MCU, use an outlet strip with a
built-in power switch. This allows the MCUs to be turned on and off
together.

2. Plug the AC adapter into a 110V AC outlet or switch on the outlet

strip to apply power to the MCU.

3. Verify that leakage current ("touch" current) is below 250 μA rms

on exposed metal surfaces.

4. If leakage is excessive, power off the system and re-verify ground

path continuity.

Set switch interface type

The following steps explain how to assign a switch interface type to

the MCU. This procedure must be done on each MCU.

MCUs can be enabled with different/mixed PBX integrations within
the same system and operate normally. However, the MCUs must
be running the same software. See Chapter 8 section Replace an
MCU
for details on the software update procedure. If you need
additional help, please contact the Polycom Customer Support for
assistance.

1. Power on the MCU with the mode switch in the

NORMAL

position.

Because the MCU does not have a switch type assigned, it will

alarm (the

ERROR

LED will flash along with

STATUS

LEDs

1

,

2

,

3

,

4

,

and

5

). This should take less than 2 minutes.

2. Power off the MCU.
3. Move the mode switch to

ADMIN

and power on the MCU. Within

15 seconds,

STATUS

LEDs

2

and

4

will light.

4. Press the

STEP

button three times.

Advertising