Chapter 4 SelfCalibration

4-3. Setting Calibration Details [Setting]

Setting the mode to execute SelfCalibration [Mode Selection]

Set the mode to execute SelfCalibration.
Procedure
1. Choose <Setting> from the SelfCalibration menu, and press

.

2. Choose <Mode Selection> from the Setting menu, and press

.

3. Select the mode to change with

or

, and press

.

4. Select “On” or “Off” with

or

.

5. Press

to exit the adjustment.

NOTE

• When

CAL1/CAL2/CAL3

is set to

On

, SelfCalibration can be executed.

• When

Other Modes

is set to

On

, the monitor color reproduction area is updated and the display in each color

mode will be adjusted as follows:

- In any mode other than CAL mode, the temperature is adjusted so that it is as close as possible to the specified

value.

- In Adobe RGB/sRGB/EBU/REC709/SMPTE-C/DCI modes, the gamut values are adjusted so that they are close

to each specified value.

Setting the calibration targets

Set the calibration targets for SelfCalibration.
Procedure
1. Choose <Setting> from the SelfCalibration menu, and press

.

2. Choose <Target> from the Setting menu, and press

.

3. Select the items to change with

or

, and press

.

The target setting menu is displayed.

4. Select the items to change with

or

, and press

.

Setting Item

Setting Range

Setting

Brightness

30 - 250 cd/m2

Set the brightness of the monitor using

the brightness value (cd/m2).

White Point

Color

coordinates

: 0.2400 - 0.4500

*1

Temperature : 4000 - 10000K (100K Unit)

*2

Set the monitor White Point using the

color coordinate or Temperature.

Color Gamut Native/User

Set whether to calibrate color gamut.

Native : Color gamut calibration is

not performed.

User : Color gamut calibration is

performed.

0.0000 - 1.0000

(For each of Red/Green/Blue)

To calibrate the monitor color

reproduction gamut, set the color

coordinates for red, green, and blue.

Gamma

1.6 - 2.7 (0.1 Unit)

*3

Set the monitor gamma values.

*1

When the color coordinates are set, the temperature changes to [User].

*2

When the temperature is set, the x and y coordinates of the white point are set automatically.

*3

These become [Fixed] when calibrating with ColorNavigator.

5. Adjust using

or

.

6. Press

to exit the adjustment.

The calibration targets are set.

Attention

• Only set the color gamut calibration when the target color range is precisely known.
• If ColorNavigator was used to calibrate the grey balance, do not change the gamma setting values.

NOTE

• The calibration targets set using ColorNavigator are set as the calibration targets for the monitor.