Teledyne Trace Oxygen Analyzer 3000TA User Manual

Page 15

Advertising
background image

Model 3000TA

Quickstart Guide

Teledyne Analytical Instruments

15

Auto Mode Zero Calibration

Press Zero to enter the zero function mode.

Use the ▲▼ arrow keys to toggle between

AUTO

and

MAN

zero settling. Stop when

AUTO

appears blinking.

Zero: Settling: AUTO

<ENT> To Begin

Press Enter to begin zeroing.

#### PPM Zero

Slope=#### ppm/s

.
.
.

#### PPM Zero

4 Left=### ppm/s

The zeroing process will automatically conclude when the output is within the acceptable

range for a good zero. Then the analyzer automatically returns to the Analyze mode.

Manual Mode Zeroing

Press Zero to enter the Zero function.

Use the ▲▼ keys to toggle between

AUTO

and

MAN

zero settling. Stop when

MAN

appears, blinking, on the display.

Zero: Settling: Man

<ENT> To Begin

Press Enter to begin the zero calibration.

#### ppm Zero

Slope=#### ppm/s

Once zero settling is completed, the instrument automatically returns to the Analyze mode.