VXI VT1433B User Manual

Page 143

Advertising
background image

Recommendations on wiring for the VT1432A/33B 4-Channel Input Connector

Allowed Connections

Differential Input Channels
Connect at VT1432A/33B end of cabling and at DUT
Recommended: shielded twisted pair

1

+ Diff 4

2

- Diff 4

5

+ Diff 3

6

- Diff 3

8

+ Diff 2

9

- Diff 2

12

+ Diff 1

13

- Diff 1

Input Channel Shielding
Connect at VT1432A/33B end of cabling ONLY

3

RFI GND/Drain Shield 4

4

RFI GND/Drain Shield 3

10

RFI GND/Drain shield 2

11

RFI GND/Drain Shield 1

Additional shielding of entire cable
GND for grounded measurements if required

14

RFI GND/Cable Shield

26

RFI GND/Cable Shield

VT1433B User's Guide

Break Out Boxes

8-7

Advertising