The aforementioned errors are reported for the following reasons:

NO_ERROR

There are no errors in the queue.

ERROR_UNKNOWN_COMMAND

An unknown command was sent to the

microprocessor.
ERROR_PRE_GT_SIZE

The value programmed in the Pre-Trigger

Points register is greater than the value

programmed in the Sample Points

register.
ERROR_RESISTANCE_OVER_RANGE

The user attempted to measure a

resistance greater than the range.

ERROR_UNSTABLE_RESISTANCE

In Auto-Range Resistance mode, the

resistance is not stable.

ERROR_UNSTABLE_VOLTAGE

In Auto-Range Voltage mode, the voltage

is not stable.

ERROR_MULTIPLE_TEST_SOURCES

The user attempted to connect several

sources on the Test Bus.

ERROR_NONVOL_READ

There was an error when trying to read

the non-volatile calibration memory.

ERROR_NONVOL_WRITE

There was an error when trying to write

to the non-volatile calibration memory.

ERROR_NONVOL_DEFAULTED

There was an error when trying to read

the default values from the non-volatile

calibration

memory.

ERROR_FLASH_BURN

Error trying to burn data in the FLASH (a

common error here would occur when

loading the firmware for U1 and issuing

the write flash commands for U20, or

vice

versa).

ERROR_INTERNAL_SOFTWARE

The microprocessor encountered an

internal software error.

ERROR_QUEUE_OVFL

There were more then five errors in the

ERROR

queue.