LF2020/LF2090 Sit-To-Stand Patient Lift Owner’s Manual

15

TROUBLESHOOTING

Symptom

Possible Cause

Action

Control box is not responding

and relay is not clicking

Batteries discharged

Battery charger defective

Batteries defective

Hand control defective

Charge batteries

Replace battery charger

Replace batteries

Replace hand control

Actuator does not run,

but relay is clicking

Actuator plug not properly

seated

Actuator defective

Control box defective

Push actuator plug firmly into

control box

Replace actuator

Replace control box

Actuator goes up or down, but

not both

Hand control defective

Control box defective

Replace hand control

Replace control box

Cannot lift patient up and

battery indicator shows no

GREEN light

Batteries discharged

Charge the batteries

Lift does not work and the battery

indicator shows full charge

Check the connection between the

actuator and the control box

Check connection between Pendant

& control box

Lift does not work and the

battery indicator shows no light

Make sure the emergency stop

button is released

Check the connection of the

batteries in control box

Stop-and-Run while lifting or

lowering patient

Check socket for pendant plug-in

Check pendant cable

Battery Indicator shows no light

when charger is connected to

power supply

Check the emergency stop button is

released.

Check charger plug for damage.

Check charger cable for damage.

Lift does not work. Battery

indicator shows full charge.

Notice “click” sound from control

box when UP or DOWN button on

hand control was pressed.

Actuator problem.

Warning: If you experience a problem with your lift and are unable to
service it yourself, contact your Graham-Field Health Products dealer.