Calibration, Equipment required – Xantrex Technology XDL 35-5T User Manual

Page 42

Advertising
background image

Calibration

Routine calibration is carried out without opening the instrument.
Allow 10 minute warm-up before commencing calibration.

Equipment Required

A 5.5 digit multimeter with better than 0.02% accuracy on dc volts and better than 0.12%
accuracy on dc current (to 5A); alternatively use a precision shunt for current measurement.

Calibration

Assign CONTROL to Output 1. To enter calibration press SHIFT, #, 99. If the instrument is a long
way out of calibration, the default calibration values should be loaded first by pressing SHIFT, #,
91. Pressing ESCAPE at anytime will abort the calibration procedure and revert to the stored
calibration values.

Having entered calibration mode, follow the table below. To increment to the next step press OK.
The calibration must be done in sequence. At the end of calibration press OK to store the
calibration. The instrument automatically sets the range and settings at each step.

Use the spin-wheel to adjust calibration; holding down the JOG SET OFF key will give x100
increments to speed up calibration. Pressing SHIFT will show the calibration value.

Repeat for Output 2.

Step/Range Settings

Adjust

for:-

Load

Detail

1, 35V 3A

0.01V 3A

10mV ± 0.5mV

DVM

o/p 0V

2, 35V 3A

0.01V 3A

0.010V on display

DVM

V readback zero

3, 35V 3A

35V 3A

35V ± 0.5mV

DVM

o/p V span

4, 35V 3A

35V 3A

35.000 on display

DVM

V readback span

5, 35V 3A

0.01V 3A

10mV ± 0.5mV

DVM

o/p 0V

6, 35V 3A

35V 3A

35V ± 0.5mV

DVM

o/p V span

7, 35V 3A

2V 0.001A

1mA ± 0.5mA

milli-ammeter

o/p A offset

8, 35V 3A

2V 0.001A

0.001 on display

milli-ammeter

A readback zero

9, 35V 500mA

2V 0.1mA

0.1mA ± 0.05mA

milli-ammeter

o/p mA offset

10, 35V 500mA

2V 0.1mA

flashing 0.0/0.1 on display

milli-ammeter

mA readback zero

change

load

11, 15V 5A

2V 4A

4A ± 0.5mA

ammeter

o/p A span

12, 15V 5A

2V 4A

4.000 on display

ammeter

A readback span

13, 15V 5A

2V 4.1A

4.100 on display

ammeter

A readback 13bit

14, 35V 500mA

2V 400mA

400mA ± 0.05mA

ammeter

o/p mA span

15, 35V 500mA

2V 400mA

400.0 on display

ammeter

mA readback span

16, 35V 500mA

2V 410mA

410.0 on display

ammeter

mA readback 13bit

Advertising
This manual is related to the following products: