Error codes, Cw-11 indicator – Ohaus CW-11 Indicator User Manual

Page 36

Advertising
background image

33

CW-11 Indicator

5.2Error Codes List

The following list describes the various error codes and which can appear on the display and the suggested

remedy.

LoBat

Is indicated when batteries are weak. Approximately 20 minutes of operating time

remain.

Error 1

Indicates an overload condition.

Error 2

Indicates an underload condition.

Error 7

EEPROM data incorrect.

Error 14 Zero exceeds ZERO% and cannot be zeroed.

Err 21

Calibration data does not match current full scale, Grad and Cal Point settings. Settings

must be restored or the Indicator must be recalibrated using the current settings.

5.3 Service Information

If the Troubleshooting section does not resolve or describe your problem, you will need to contact an authorized
Ohaus Service Agent. For Service assistance in the United States, please call Aftermarket, Ohaus Corporation toll-
free at (800) 526-0659. An Ohaus Product Service Specialist will be available to help you.

5.4 Technical Data

Materials
Housing stainless-steel
Keypad/display overlay polyester

Ambient conditions
The technical data is valid under the following ambient conditions:
Ambient temperature

-10°C to 40C/ 14°F to113°F

Relative humidity

10%.......95%, noncondensing

Height above sea level

up to 4000m

Operability is assured at ambient temperatures between 5 and 40° C.