4 Mounting

SOLITRAC 31 • 4 … 20 mA/HART - four-wire

43387-EN-131119

To measure the level in very high vessels, multiple instruments can be

cascaded.
Cascading means that two or several instruments are connected

which can together cover a longer measuring range.
You can find the exact number of Slaves in the Safety Manual.

L1

L2

h

1

Fig. 4: Installation position - cascading arrangement
h Summed measuring range

L Measurement length (L1, L2)

1 Red marking line for designating the measuring range

Here, one instrument acts as a Master and all other instruments

operate as Summation Slaves. The pulse rates of all instruments

are summed in the Master instrument and converted into a common

signal.
If several sensors are cascaded, the measuring ranges of the indi-

vidual detectors must directly join each other. The detectors must also

slightly overlap.
Make sure that the red marking lines directly join the measuring range

of the next SOLITRAC 31.
Mount the SOLITRAC 31 in such a way that the detector tube is

directly in the radiated area of the source container. Mount the SOLI-

TRAC 31 preferably side by side and make sure that no detector tube

is hidden by another sensor.

Summation