Goldeye Technical Manual V2.5.0

61

Example

Start the correction data selection once and activate the best data set based on
the internal statistics value.

>

k=0[CR]

S=0A
>

Goldeye xy-008

If

k is not set to 0 the automatic search starts at approx. each k * 256

images. At a frame rate of e.g. 50 images per second the chronological resolu-
tion of the

k value is approx. 5.1 seconds.

Goldeye xy-032

If

k is not set to 0 the automatic search starts at approx. each k * 256

images. At a frame rate of e.g. 30 images per second the chronological resolu-
tion of the

k value is approx. 8.53 seconds.

Example

Goldeye xy-008

The automatic correction data selection shall always be started approx. every
1200 seconds (20 minutes): 1200 s / 5.1 s per count ≈ 235 = 0xEB.

Goldeye xy-032

The automatic correction data selection shall always be started approx. every
1200 seconds (20 minutes): 1200 s / 8.53 s per count ≈ 140 = 0x8C.

>

k=EB [CR](example Goldeye xy-008)

S=0A(...at once)
>

S=0B(...after each 235 * 256 images...)

(...)

Select a correction data set (S)

All Goldeye xy-... cameras are equipped with multiple correction data sets for
different operation conditions. The image quality can be improved by activating
another correction data set if the environmental conditions have changed.

GigE feature name (CameraSpecialFeatures)

Feature Visibility Description

AutoCalibrateOnce

Beginner

Start the automatic calibration
once.

(

k=0 command)

The processing of this com-
mand can take several seconds,
depending on the current
frame rate and the number of
correction data sets available.

Table 28: GigE feature: AutoCalibrateOnce