The M305 blackbody calibration source design achieves significant weight and size reduction
over the M300 with only minor sacrifices in specification and performance, making it an ideal

unit where portability is required or space is limited. The spherical cavity yields an emissivity of
0.995 over the temperature range of 100 °C to 1000 °C with an aperture of 25 mm. An integrally
mounted PID controller completes this versatile calibrator.

2.2 Technical Data

Temperature Range:

100 °C to 1000 °C or 210 °F to 1832 °F

Accuracy:

±0.25% of reading ±1 C when using Calibration Certificate

correction factors

Temperature Resolution:

0.1 C

Stability:

<1 C per 8 hour period

Source Non-Uniformity:

±0.2 C typical at T<230 °C, ±1 °C typical at T>230 °C

Heated Cavity Shape:

Spherical

Aperture Diameter:

25.4 mm (1.00”)

Emissivity (calculated from

cavity shape):

0.9995 ± 0.0005

Effective emissivity:

1.00 @ 8-14 µm T<230 °C, 1.00 @ 0.7-1.8 µm T>230 °C

Method of Control:

Digital self-tuning PID controller

Warm-up Time:

60 minutes from ambient to 700 °C

Operating Ambient

Temperature:

0 C to 44 C or 32 F to 110 F

Cooling:

Fan cooled, air inlet on rear panel

Operating Humidity:

90% RH max, non-condensing

Power Requirements:

[ ]120VAC [ ]208VAC [ ]230/240VAC [ ]60Hz [ ]50Hz

Dimensions:

270mm H x 430mm W x 370mm D

Weight:

55 lb. (25 kg)