LumaSense Technologies M340 User Manual

Page 25

Advertising
background image

M340 Manual

Appendix B: Calibration Methods 25

To force the radiance output of a thermometrically calibrated blackbody to desired
temperature, simply invert the sign of the ERROR DEG.C and add this number to the blackbody
controller setpoint. Again, note that this only applies at ambient and background temperatures

of 23 °C.

Calibrating fixed emissivity pyrometers

To use this blackbody to calibrate pyrometers (infrared thermometers) that are factory set at
0.95 emissivity, use the table below to find the proper M340 controller setpoint temperature
that will provide the correct output radiance for e=0.95. Please review all notes provided below
the table.