Fig.12.4 Part of a calibration data print-out

12.3.3 Checking and Changing Optical Filter Parameters

Before starting calibration it is always wise to check that information

about the installed optical filters has been “entered” correctly in the

active set-up of the Monitor. If the incorrect UA number is

“entered” for the optical filter in a particular position, the incorrect

optical filter factors will be used in the calculation of the calibration

factors, and calibration will therefore be useless.

The UA number can only be changed using the push-buttons on the

front panel of the monitor.

1. Press SET-UP S3 S3.

The following text appears on the screen display: