Additel 221A Multifunction Temperature Calibrator User Manual User Manual

Page 70

Advertising
background image

 

 

  ________________________________________________________________________________________________________________  

63 

 

Figure 10.8 AsCal Calibration Mode 

The Calibration Mode page will appear if the UUT is indicator or recorder 
(as shown in Figure 10.8).
About Reverse Calibration
Adjust the output signal of the calibrator until the reading of the UUT 
displays a nominal set point, then record the reading from the calibrator. 
The calibrator will calculate the UUT's theoretical reading accordingly. 
In most cases, the reverse calibration is a very practical method, which 
eliminates the trouble of manual recording. 
Pressure indictors/recorders are also suitable for reverse calibration. You 
can apply a pressure to read the pressure gauge then record the reading 
from the pressure module (calibrator). 
About Forward Calibration
The forward calibration is traditional, simple ways to calibrate. (To make 
the calibrator output a nominal set point signals, and then record the 
UUT's actual readings.) 

 