Dwyer AP2 User Manual

Dwyer instruments, inc, Model ap2 wireless thermo-anemometer probe for uhh, Bulletin te-ap2

Advertising
background image

Model AP2 Wireless Thermo-Anemometer Probe for UHH

Specifications - Installation and Operating Instructions

Bulletin TE-AP2

DWYER INSTRUMENTS, INC.

Phone: 219/879-8000

www.dwyer-inst.com

P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.

Fax: 219/872-9057

e-mail: [email protected]

SPECIFICATIONS
Service:
Dry, clean air.
Temperature Limits:

Process: -20 to 200°F (-29 to 93°C);

Ambient: -20 to 140°F (-29 to 60°C);

Battery Charging: 32 to 113°F (0 to 45°C).

Range:

Air Velocity: 0 to 6000 FPM (0 to 30 m/s);

Volumetric Air: 999,999 in selected flow units;
Temperature: -40 to 212°F (-40 to 100°C).

Accuracy:

Air Velocity: ±3% FS within temperature range of 40 to 90°F (4 to 32°C);

Temperature: ±0.5°F (±0.28°C).

Response Time: 1 s.
Probe Length: 8˝ (203 mm) insertion.
Power Requirements: 3.7 V YT562447 Lithium ion battery, installed functional,
user replaceable.
Maximum Wireless Distance: 50´ (15 m).
Handle Enclosure: Thermoplastic elastomer over polycarbonate.
Supplied With: Wrist strap.
Weight: 11.2 oz (331.22 g).
Agency Approvals: CE (not while charging), RoHS, FCC compliant.

The Model AP2 Wireless Thermo-Anemometer Probe measures air velocity, air
flow, and temperature when combined with the Model UHH Universal Handheld. In
order to prevent damage to the sensing element, the probe has a protective sleeve
that slides over the sensor when it is not in use. Markings on the probe allow users
to know the insertion depth to obtain better accuracy when traversing the duct.

Wireless probes can take measurements up to 50 feet away from the UHH. A bi-
color LED flashes on the handle of the probe to indicate communication status with
the UHH. The battery is rechargeable via the mini-USB connector on the bottom of
the probe.

CHARGING BATTERY

When the charge of the battery is almost used up, the LED on the handle of the
wireless probe will turn solid red. There is approximately 5 minutes of battery life
left at this point. If the probe is not already paired to the UHH, it will not pair while
in low battery condition.

Step 1: Open the USB cover on the bottom of the probe handle.
Step 2: Plug the mini-USB connector end of the cable into the probe handle.
Step 3: Plug the USB connector end of the cable into the port on the charger or
PC.
Step 4: Plug the charger into either an electrical outlet or car charging port (LED
on the charger and the handle should both light up).
Step 5: The LED on the handle will turn off when fully charged.
Step 6: Remove charger from electrical outlet.
Step 7: Remove USB connectors from the handle and the charger.
Step 8: Replace the USB cover on the bottom of the probe handle.

1 IN

-

-

2

-

-

3

-

-

-

4

-

5

-

-

-

6

-

-

7

-

8

5/16

[7.94]

15-3/8

[390.68]

6-3/64

[153.63]

1-41/64

[41.62]

If desired, can be operated with USB cables less than 3 m in
length when connected to the charger or PC.

NOTICE

It is required prior to the initial usage to charge the battery for
12 hours.

NOTICE

Lithium ion polymer batteries are very volatile and can cause a
fire if punctured or severely damaged. Only use a Dwyer

Instruments, Inc. approved charging device in a well ventilated area away from any
flammable materials or gases. Do not incinerate the battery. Only charge between
32 to 113°F (0 to 45°C).

WARNING

This device complies with Industry Canada license-exempt RSS
standard(s). Operation is subject to the following two conditions:

(1) This device may not cause harmful interference. and (2) this device must accept
any interference received, including interference that may cause undesired
operation.

Cet appareil est conforme à des règlements d'Industrie Canada exempts de licence
standard RSS (s). Son fonctionnement est soumis aux deux conditions suivantes:
(1) Ce dispositif ne doit pas causer d'interférences nuisibles, et (2) cet appareil doit
accepter toute interférence reçue, y compris les interférences pouvant entraîner un
fonctionnement indésirable.

This Class B digital apparatus complies with Canadian ICES-003.
Cet appareil numériqué de la classe B est conformé à la norme NMB-003 du
Canada.

NOTICE

The side of the probe with the dot should
face the flow.

NOTICE

Advertising