Dwyer VP2 User Manual

Dwyer instruments, inc, Bulletin te-vp2

Advertising
background image

Model VP2 Wireless 100 mm Vane Thermo-Anemometer Probe for UHH

Specifications - Installation and Operating Instructions

Bulletin TE-VP2

DWYER INSTRUMENTS, INC.

Phone: 219/879-8000

www.dwyer-inst.com

P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.

Fax: 219/872-9057

e-mail: [email protected]

SPECIFICATIONS
Service: 
Clean air.
Temperature Limits:

Process: -4 to 140°F (-20 to 60°C);

Ambient: 5 to 125°F (-15 to 51°C).

Range:

Air Velocity: 50 to 5000 fpm (0.25 to 25 m/s);

Temperature: -22 to 140°F (-30 to 60°C);
Relative Humidity: 0 to 100% RH;

Air Volume: 999,999 in selected units.

Accuracy:

Air Velocity: 0.25 to 10 m/s: ±1.5% of reading ±20 fpm (±0.1 m/s);

10 to 20 m/s: 1.5% of reading ±40 fpm (±0.2 m/s); 20 to 25 m/s:
±1.5% of reading ±60 fpm (±0.3 m/s);

Temperature: ±0.54°F @ 77°F (±0.3°C @ 25°C);
Relative Humidity: ±2% @ 77°F (25°C) (10 to 90% RH); ±4%

(0 to 10% and 90 to 100%).

Response Time:

Air Velocity: 1 s;

Temperature: 1.5 s;
Relative Humidity: 1.5 s;

Air Volume: 1 s.

Probe Length: 8˝ (203 mm) insertion.
Power Requirements: 3.7 V YT562447 Lithium ion battery, installed functional,
user replaceable.
Battery Charging Limits: 32 to 113°F (0 to 45°C).
Maximum Wireless Distance: 50´ (15 m).
Vane Material: Anondized aluminum.
Handle Enclosure: Thermoplastic elastomer over polycarbonate.
Supplied With: Wrist strap.
Weight: 13.6 oz (402.20 g).
Agency Approvals: CE (not while charging), RoHS, FCC compliant.

The Model VP2 Wireless 100 mm Vane Thermo-Anemometer Probe measures
air velocity, air flow, humidity, and temperature when combined with the Model
UHH Universal Handheld. By having a larger diameter, the rotating vane is able to
measure velocities down to 50 fpm or 0.25 m/s. An arrow is molded into the vane
housing to depict the flow direction.

CHARGING BATTERY

When the charge of the battery is almost used up, the LED on the handle of the
wireless probe will turn solid red. There is approximately 5 minutes of battery life
left at this point. If the probe is not already paired to the UHH, it will not pair while
in low battery condition.

Step 1: Open the USB cover on the bottom of the probe handle.
Step 2: Plug the mini-USB connector end of the cable into the probe handle.
Step 3: Plug the USB connector end of the cable into the port on the charger or
PC.
Step 4: Plug the charger into either an electrical outlet or car charging port (LED
on the charger and the handle should both light up).
Step 5: The LED on the handle will turn off when fully charged.
Step 6: Remove charger from electrical outlet.
Step 7: Remove USB connectors from the handle and the charger.
Step 8: Replace the USB cover on the bottom of the probe handle.

It is required prior to the initial usage to charge the battery for
12 hours.

NOTICE

Lithium ion polymer batteries are very volatile and can cause a
fire if punctured or severely damaged. Only use a Dwyer

Instruments, Inc. approved charging device in a well ventilated area away from any
flammable materials or gases. Do not incinerate the battery. Only charge between
32 to 113°F (0 to 45°C).

WARNING

4-13/64

[106.65]

16-25/32

[426.08]

1-5/32

[29.31]

2

[50.80]

If desired, can be operated with USB cables less than 3 m in
length when connected to the charger or PC.

NOTICE

This device complies with Industry Canada license-exempt RSS
standard(s). Operation is subject to the following two conditions:

(1) This device may not cause harmful interference. and (2) this device must accept
any interference received, including interference that may cause undesired
operation.

Cet appareil est conforme à des règlements d'Industrie Canada exempts de licence
standard RSS (s). Son fonctionnement est soumis aux deux conditions suivantes:
(1) Ce dispositif ne doit pas causer d'interférences nuisibles, et (2) cet appareil doit
accepter toute interférence reçue, y compris les interférences pouvant entraîner un
fonctionnement indésirable.

This Class B digital apparatus complies with Canadian ICES-003.
Cet appareil numériqué de la classe B est conformé à la norme NMB-003 du
Canada.

NOTICE

Advertising