Chapter 15 _________________________________________________Calibrationd and Adjustment

VAISALA _______________________________________________________________________ 83

6.

Turn the service switch 4 to ON-position (adjustment enabled),
see page 18, Figure 4, DMP501 back panel, service switches.

7.

Open

MENU

: press .

8.

Select ►

Settings

, press .

9.

Select ►

Adjustments

, press .

10.

Select ►

Dewpoint adjustments

, press .

11.

Press SET. Give the reference value (

Ref.

) by using the

numeric buttons. Then, give the measured value (

Meas.

). Press

OK

.

12.

Select with the arrow button the next reference point and press

SET

.

13.

Continue giving the references.

14.

Make the adjustment by pressing

Adjust

.

15.

The adjustment result is shown.

16.

Press

EXIT

.

17.

Turn the service switch 4 to OFF-position (adjustment disabled).

The system calculates a 3rd degree correction polynomial fitted to the
adjustment data.