Figure 27, Sampling system upside down, Warning – Vaisala DM70 User Manual

Page 60

Advertising
background image

USER'S GUIDE____________________________________________________________________

2.

Turn the sampling system upside down. See Figure 27 below.

0403-050

Figure 27

Sampling System Upside Down

The following numbers refer to Figure 27 above.

1 = Battery
2 = Battery

wires

3

=

Nut of the sample cell

3.

Detach the battery wires. Remove the old battery by pulling it
up. The battery is attached to its place with double-sided tape.

NOTE

The used battery must be disposed of according to the local laws and
regulations.

4.

Attach the double-sided tapes provided with the battery to the
new battery enclosure. Place them onto the side without text.
For reference, see the old battery.

WARNING

Batteries are electrochemically live at all times. Do not short-circuit
them.

5.

Reconnect the wires as follows:
Red pole: BA + wire
Black pole: BA - wire

6.

Press the new battery onto the tapes.

58 __________________________________________________________________ M010091EN-F