Chapter 5 ______________________________ Operating WMT700 with WS425 and SDI-12 profiles

VAISALA _______________________________________________________________________ 59

WS425 A/B WAT11 Commands

When the WS425 A/B WAT11 profile is selected, you can poll data
using the following command:

<esc><id>

where
<esc> = Escape character ASCII 27H
<id> = The WMT700 ID, for example, A

The response is the following:

<stx><id><spd><dir>

where
<stx> = Start of text character (1 digit)
<id> = The WMT700 identification character, for example, A (one

digit).

<spd> = Wind speed (in meters per second) multiplied by 10. For

example, 045 is 4.5 meters per second (three digits).

<dir> = Wind direction with two octal numbers for 6-bit binary data,

for example, 73 corresponds to (7*8+3)/64*360 = 332
degrees.

Missing Readings

The WS425 A/B WAT11 message reports missing data as slashes (/////).