Checking the output power, Table 3-2 – Agilent Technologies E8257D/67D User Manual

Page 29

Advertising
background image

Chapter 3

25

Operation Verification

Checking the Output Power

Checking the Output Power

This test verifies the CW output power from the signal generator. If this test fails, refer to

“Problems with

Output Power” on page 32

for further instructions.

Table 3-2

shows the recommended equipment for use with the E8257D signal generator when performing

this test.

Table 3-3

shows the recommended equipment for use with the E8267D signal generator when performing

this test.

Table 3-2

Recommended Equipment for the E8257D

E8257D

Option

520

E8257D

Option

540

E8257D

Option

550

E8257D

Option

567

Test Equipment

Recommended Agilent

Test Equipment

Power Meter

Agilent E4418B E-Series power meter
or Agilent E4419A/B E-Series power
meter

Power Sensor,
Input: Type-N
(m)

Agilent E9304A E-Series power sensor
or Agilent 8482A power sensor

Power Sensor,
Input: APC 3.5
(m)

Agilent E4413A E-Series power sensor

Power Sensor,
Input: 2.4 mm
(m)

Agilent 8487A Option H84 power
sensor (Option H84 adds a Standards
Lab Calibration)

V-Band Power
Sensor

Agilent V8446A V-Band Power Sensor

Adapter,
2.4 mm (f) to
Type-N (f)

Agilent 11903B 2.4 mm (f) to
Type-N (f) adapter

Adapter,
2.4 mm (f) to
(f)

Agilent 11900B 2.4 mm (f) to (f) adapter

Advertising