4 application parameters, Operating manual – Eppendorf 5810 R Centrifuge User Manual
Page 72
Centrifuge 5804/5804 R/5810/5810 R — Operating manual
72
The noise level was measured according to DIN EN ISO 3745 frontally in a sound measuring
room with accuracy class 1 at a distance of 1 m from the device and at lab bench height.
10.4
Application parameters
Weight without rotor
5804
55 kg (121 lb.)
5804 R
80 kg (176 lb.)
5810
68 kg (150 lb.)
5810 R
99 kg (218 lb.)
Rotor
A-4-81
(4 × 500 mL)
A-4-44
(4 × 100 mL)
F-34-6-38
(6 × 85 mL)
Noise level
5804
-
< 67 dB(A)
< 51 dB(A)
5804 R
-
< 56 dB(A)
< 58 dB(A)
5810
< 65 dB(A)
< 65 dB(A)
< 53 dB(A)
5810 R
< 56 dB(A)
< 56 dB(A)
< 59 dB(A)
Runtime:
1 to 99 min, adjustable in 1 min increments.
infinite ( )
Temperature (only 5804 R/
5810 R):
-9°C to 40°C
Relative centrifugal force
(RCF or rcf):
10 to 20,913 × g
adjustable up to 3,000 × g in 10 × g increments, thereafter in
100 × g increments.
Speed:
200 to 14,000 rpm,
adjustable up to 5,000 rpm<Sup/> in 10 rpm<Sup/>,
afterwards in 100 rpm increments.
Max. load:
5804/5804 R:
5810/5810 R:
4 × 250 mL
4 × 750 mL
Max. kinetic energy:
5804/5810:
5804 R:
5810 R:
19,000 Nm (11,000 rpm)
19,000 Nm (11,000 rpm)
23,000 Nm (12,000 rpm)
Test log mandatory in Germany:
Yes
Permitted density of the centrifugate
(at max. g-force/rpm and max. load):
1.2 g/mL
Standardized interface (optional)
RS 232 C
EN
Operating manual