Control Company 4332 ENVIRO-METER User Manual

Enviro-meter™ instructions

Advertising
background image

Traceable

®

is a registered trademark of Control Company ©2008 Control Company. 92-4332-00 Rev. 2 052308


FEATURES

4 in 1 professional measuring instrument: Anemometer, Hygrometer,
Thermometer, and Light meter.

Tiny bone shape with lightweight and small size case design are suitable for
handling with one hand.

Wristlet design provides extra protection to the instrument especially for user
one hand operation.

Low-friction ball bearing mounted wheel design provides high accuracy at high
and low air velocity.

Exclusive photo diode and color correction filter light sensor, spectrum meets
C.I.E. photopic.

High precision thin-film capacitance humidity sensor with fast response to the
humidity changes.

Standard type K (NiCr-NiAl) thermocouple input jack suitable for all kinds of type
K probe.

Built- in microprocessor circuit assures excellent performance and accuracy.

Concise and compact buttons arrangement, easy operation.

Memorize the maximum and minimum value with recall.

°C/°F selectable by pressing button on the front panel.

Lux-Foot-candles selectable by pressing button on the front panel.

Air velocity measuring units selectable by pressing button on the front panel for
five kinds of units.

Multi channel display for relative humidity and temperature measured values or
air velocity and temperature measured values at the same time.

Zero button design makes light meter calibration.

Hold function to freeze the current reading value.

SPECIFICATIONS

Display

1

/

3

-inch LCD display

Measurement

Anemometer, Humidity, Temperature, Light

Operating Humidity

Max. 80% RH.

Operating Temperature 0 to 50 °C (32 to 122 °F)
Over Input Display

Indication of “- - - -”

Power Supply

006P DC 9V battery (Heavy duty type)

Power Consumption

Approx. DC 6.2 mA

Weight

5¾ ounces (battery included)

Dimensions

7 x 2½ x 1¼ inch.


3-1

Power Button

3-2

Hold Button

3-3

Max./Min. Button

3-4

Unit/Zero Button

3-5

°C/°F Button
Lux/Fc Button

3-6

Function Button

3-7

Air Flow Sensor

3-8

Thermocouple Input Sensor

3-9

R.H. Sensor

3-10 Light Sensor
3-11 LCD display
3-12 Battery Compartment/Cover
3-13 Wristlet

ft/min: feet per minute MPH: miles per hour m/s: meters per second knots: nautical
miles per hour km/h: kilometers per hour Fc: foot candles








Measurement

Range

Resolution

ft/min

80 to 5910 ft/min

1 ft/min

m/s

0.4 to 30.0 m/s

0.1 m/s

km.h

1.4 to 108.0 km/h

0.1 km/h

MPH

0.9 to 67.0 mile/hr

0.1 MPH

knots

0.8 to 58.3 knots

0.1 knots

32 to 122°F

0.1 °F

Air Velocity

Temperature

(thermistor)

0 to 50°C

0.1°C

Measurement

Range

Resolution

% RH

10 to 95% RH

0.1% RH

32 to 122 °F

0.1 °F

Humidity

Temperature

(thermistor)

0 to 50 °C

0.1 °C

0 to 2,200 Lux

1 Lux

Lux

1,800 to 20,000

Lux

10 Lux

0 to 204.0 Fc

0.1 Fc

Light (auto range)

Foot Candles

170 to 2,000 Fc

1 Fc

–148 to 2372°F

0.1°F

Temperature (Type-K)

–100 to 1300°F

0.1°C

Measurement

Range

Accuracy

80 to 5910 ft/min

0.4 to 30.0 m/s

1.4 to 108.0 km/h

0.9 to 67.0 mile/h

0.8 to 58.3 knots

≤ 20 m/s : ± 3% F.S.

> 20 m/s : ± 4% F.S.

32 to 122°F

±2.5°F

Air Velocity

0 to 50°C

±1.2°C

10 to 95% RH

< 70% RH: ±4% RH

≥ 70% RH: ± (4% or

reading + 1.2% RH)

32 to 122 °F

±2.5°F

Humidity

0 to 50 °C

±1.2°C

0 to 20,000 Lux

Light

0 to 2,000 Fc

± 5% of reading + 8

digits

–148 to 2372°F

± (1% of reading +2°F

Temperature (Type-K)

–100 to 1300°F

± (1% of reading + 1°C)






ENVIRO-METER™ INSTRUCTIONS

Advertising