DESCRIPTION

Traceable

®

Logger-Trac™ is the revolutionary, multi-use,

user programmable, electronic temperature data recorder

specifically designed for measuring environmental tem-

peratures during the transportation of refrigerated vaccines,

biological material, chemicals, pharmaceuticals, and for

documenting the transportation of perishable commodities.

FIGURE 1

LCD DISPLAY

CURRENT TEMPERATURE – The current temperature

reading.

MIN/MAX RECORDINGS – The minimum and maximum

temperatures that have occurred since the device was

initialized.

OVER ALARM COUNTER – Indicates the count of

temperature logged readings which were outside the alarm

temperature limits.

ALARM WARNING – Symbol which becomes active once

a single reading is recorded outside the alarm temperature

limits.

TIME OVER ALARM – Indicates the cumulative time the

temperature has been outside the alarm temperature limits.

START/STOP BUTTON – The button is used to start the

device logging once it has been initialized. Hold button

down for 5 seconds to activate logging. If the Stop key

feature is enabled, pressing and holding the button will stop

logging preemptively. Note: if the thermometer is stopped

using the Stop button, the thermometer must be reinitialized

using the MaxiThermal software before logging may be

restarted.

QUARANTINE WARNING – When using the quarantine

alarm feature, the symbol will become active once the

quantity of temperatures logged outside the alarm limits

reaches the quarantine programmed trigger value.

Quarantine function is disabled in factory settings.

QUARANTINE TRIGGER VALUE – The number of

temperature recordings outside the alarm limits required to

activate the Quarantine warning symbol.

ALARM TRIGGERED – Indicates the temperature is in an

active alarm state.

RUN TIME (RECORDING TIME) – Indicates the amount of

time passed since the thermometer logging was initialized/

started.
PRE-PROGRAMMED SETTING

All units are pre-programmed at factory with the following

settings:

• Temperature units: Celsius

• Alarm setting – Minimum: 2°C Maximum: 8°C

• Recording duration: 30 days

• Recording interval: 5 minutes 33 seconds

• User Stop Key Enable: enabled

• Quarantine Warning: disabled
PROGRAMMABLE USER SETTINGS

Through MaxiThermal software (not included) and USB

interface (not included), user is able to program the unit

to their specific requirements. Programmable options are

listed as follows:

Description – Each Logger-Trac™ device may be

programmed with a unique description. This allows the

user to identify at the time of download where the device

was used.

Time zone – Set time on unit to a specific time zone.

Auto Adjust Daylight Time enable – The internal clock

will automatically adjust when daylight savings time shift

occurs.

Temperature Measurement Units (Celsius/Fahren-

heit) – The temperature data will display on the LCD in

the selected measurement units.

Start delay – When this feature is used, a time delay

occurs before data recording is started after the user

presses the START button.

Recording duration/Sample Interval – User can either

select the data recording interval or the total recording

duration. Note the parameters are dependent upon

each other so changing one field also changes the other.

Memory Configuration – Once data memory is full, the

device may be set to either stop recording (“Record to

End of Memory” option) or continue recording by over-

writing the oldest data (“Continuous Memory” option).

Alarm setting – Maximum and minimum alarm tem-

perature boundaries can be set. Setting resolution is 1°.

Quarantine value – The quarantine feature allows

a user to program fixed values which designate how

many temperature readings can be recorded outside

the defined alarm limits before the “QUARANTINE” LCD

symbol should become active.

No alarm – Disable the alarm with this option.