6.1.4. Connecting Controllers to a Communications Host PC

6.1.4.1. Overview of Networking Multiple Protocol Plus Controllers

The host PC must have its assigned RS-232 COM port connected to a RS-422/RS-485 bi-
directional adapter, for example a “B&B Electronics” adapter. These P/Ns provide the adapter,
power supply and PC-to-adapter serial cable. The following wiring description refers specifically
to the B&B device, but may be applied to acceptable substitute devices.

The Protocol Plus controller’s optional communication card should be configured for RS-422A 4-
wire connection. Set the dipswitches and jumpers on the adapter to match the communication
from the host PC RS-232 COM port to the adapter. Wiring from the B&B adapter’s RS-422A
RS-485 terminals is daisy-chained to each of the Protocol Plus controllers. Connect the power
supply pigtail leads from the adapter wall mount transformer into the +12V and GND on the
adapter. Plug the transformer into a standard 120V / 60 Hz supply outlet.

Note that the communication ready ovens are supplied with two DB9 network ports;
“Communication In” and “Communication Out.”

For Shorter Networking Runs
Standard serial communication rated cables are available in a variety of lengths. Contact
Despatch Service for more information (see Section 1.2).

For Longer Network Runs
RS-422A/RS-485A cables typically consist of three individually shielded twisted pairs:

One pair transmits data from the host PC to the ovens

One pair transmits data from the ovens to the host PC

One conductor from the third pair is the signal common

The remaining conductor of the third pair is unused.


Cable such as Belden type 9843 should be used.

The network requires a cable breakout junction box near each oven. 120 ohm terminating
resistors must be installed at each end of the RS-422/RS-485 network (see Figure 48 for a


If the ovens purchased were not supplied with communication
options, order additional components (add-on Protocol Plus
communication board, internal cabling, DB9 external cable
connections) to make the ovens communication-ready. Order
components required to network two ovens to a host PC.

Contact Despatch Service for more information (see Section
1.2).