2 replace mains fuse, Replace mains fuse, Danger – INFICON Helium Leak Detector Modul1000 User Manual

Page 83

Advertising
background image

Maintenance tasks

7-11

jinb

80

e1

-h

Op

er

at

in

g M

a

nua

l

(1

01

2)

7.7.2

Replace mains fuse

Required Tools

Screwdriver size 1

Required material

Safety fuse 2xT 6.3A

The main fuses of the Modul1000 are housed in the mains switch of the equipment.

For replacing the main fuses, open the cover of the fuse insert using a screwdriver
For this, insert the screwdriver into the respective groove and open the cover.
Refer to

Fig. 7-5

.

STOP

Danger

Danger of fatal injuries due to electric shock.

Before replacing the fuse, disconnect the power plug of the Modul1000.
Only insert fuses with the specified values in the Modul1000.

Fig. 7-5 Fuse replacement

Item Description

Item Description

1

Cover

3

Fuse insert

2

Safety fuse T 6.3A

4

Groove openings

1

2

3

4

Advertising