INFICON D-TEK CO2 Refrigerant Leak Detector User Manual

Page 10

Advertising
background image

10

3) After warm up
sequence the green,
the last yellow and
the amber charging
LEDs flash.

3a) Battery needs
recharging.

3a) Charge battery for 10-12
hours.

3b) Power stick has
failed.

3b) Replace power stick with
part #712-700-G1.

4) Very erratic, goes
off when bumped
slightly.

4a) CO

2

in

background may be
high or changing
quickly

4a) Take steps to reduce
concentration or to limit
changes in background
(e.g., operator wearing a
mask).

5) Pump is not
working.

5a) Pump has failed.

5a) You should hear the
pump running after the
D-TEK CO2 has completed
its warm up sequence and is
held up to your ear. VERIFY
it is not a low battery
condition (see #3). If not,
contact INFICON.

6) Unit does not
power up. Hi/Lo
indicators flash in a
rapid, alternating
manner when AC
adapter is
connected.

6a) Power
stick/battery has
been connected in
reverse.

6a) Access power stick.
Disconnect and re-connect
paying attention to the key
connector.

Problem

Cause

Remedy