PM-700 SERIES PLATING MONITOR

55

Header

(2 bytes)

Instruction Code

(1 byte)

Message
Length

(1 byte)

Command Code
or Message

Checksum

(1 byte)

FFh

FEh

30h 1h

40h 8Eh

4.8.4.4 Remote Monitoring

4.8.4.4.1

Upload Machine Status (Instruction 40h)

This operation outputs all relevant machine variables, flag registers, and I/O
registers to the host computer. The variables are output in the following order:

Table 10 Upload Machine Status Codes

Number

Position

Description

Length

(Bytes)

1

1

Rate scale factor x 256

3

2

4

Thickness scale factor x 256

2

3

3

7

Current scale factor x 256

2

3

4

10

Rate decimal position

1

5

11

Thickness decimal position

1

6

12

Current decimal position

1

7

13

Monitor status register

1

8 14 Failure

register

1

9 15 Flag

register

1

10 16 Mode

register

1

11

17

RS-232 error register

1

12

18

Output #1 status register

1

13

19

Output #2 status register

1

Total 19