2 technical data – INFICON SKY CDG045D 0.05 ... 1000Torr / mbar User Manual

Page 9

Advertising
background image

tina51e1-c (2011-12)

9

2 Technical

Data

For further technical data for gauges with DeviceNet and

Profibus interface

→  [5] and [6].

Measurement range

→ "Validity"

Accuracy

1)

0.15% of reading

Temperature effect on zero

0.1 / 0.25 / 0.5 F.S.
1 … 1100 F.S.


0.0050% F.S./ °C
0.0025% F.S./ °C

Temperature effect on span

0.01% of reading / °C

Resolution

0.003% F.S.

Gas type dependence

none


Output signal analog
(measurement signal)

Measurement range

0 … +10 V

Voltage range

–5 … +10.24 V
(limited to +10.24 V)

Relationship voltage-pressure

linear

Output impedance

0

Ω (short-circuit proof)

Loaded impedance

>10 k

Ω

Response time

2)

≥0.25 Torr/mbar (F.S.)
0.1 Torr/mbar (F.S.)


30 ms
130 ms


Identification

Resistance R

Ident

Voltage


13.2 k

Ω referenced to

supply common
≤5 V

1)

Non-linearity, hysteresis, repeatability at 25 °C ambient operating

temperature without temperature effects after operation of 2 h.

2)

Increase 10 … 90 % F.S.R.