2 technical data – INFICON SKY CDG045D 0.05 ... 1000Torr / mbar User Manual
Page 9
tina51e1-c (2011-12)
9
2 Technical
Data
For further technical data for gauges with DeviceNet and
Profibus interface
→ [5] and [6].
Measurement range
→ "Validity"
Accuracy
1)
0.15% of reading
Temperature effect on zero
0.1 / 0.25 / 0.5 F.S.
1 … 1100 F.S.
0.0050% F.S./ °C
0.0025% F.S./ °C
Temperature effect on span
0.01% of reading / °C
Resolution
0.003% F.S.
Gas type dependence
none
Output signal analog
(measurement signal)
Measurement range
0 … +10 V
Voltage range
–5 … +10.24 V
(limited to +10.24 V)
Relationship voltage-pressure
linear
Output impedance
0
Ω (short-circuit proof)
Loaded impedance
>10 k
Ω
Response time
2)
≥0.25 Torr/mbar (F.S.)
0.1 Torr/mbar (F.S.)
30 ms
130 ms
Identification
Resistance R
Ident
Voltage
13.2 k
Ω referenced to
supply common
≤5 V
1)
Non-linearity, hysteresis, repeatability at 25 °C ambient operating
temperature without temperature effects after operation of 2 h.
2)
Increase 10 … 90 % F.S.R.