Current consumption

<500 mA
(max. starting current)

Power consumption
(depending on supply voltage)


≤1 W

Fuse required

2)

1 AT (slow), automatic reset
(Polyfuse)

The gauge is protected against reverse polarity of the supply
voltage.

Electrical connection

15 poles D-Sub, male

Sensor cable

without switching functions
with switching functions


5 poles plus shielding
9 poles plus shielding

Cable length

≤100 m (0.14 mm² conductor)

For longer cables, larger conductor cross-sections are required
(R

cable

1.0

Ω).

Grounding concept

Vacuum flange - signal common

→ "Electrical Connection"

Supply common - signal common conducted separately; for dif-

ferential measurement (10

Ω)

Materials exposed to vacuum

Flange, tube
Sensor and diaphragm
Sensor–diaphragm connection
Ceramics–metal connection

stainless steel AISI 316L
ceramics (Al

2

O

3

≥99.5%)

glass ceramics solder
AgTiCu hard solder, Vacon 70
(28% Ni, 23% Co, 49% Fe)

Internal volume

≤3.6 cm

3

Admissible pressure (absolute)

≥200 Torr/mbar (F.S.)
1 … 100 Torr/mbar (F.S.)
0.1 / 0.25 Torr/mbar (F.S.)


3 bar
2 bar
1.3 bar

Bursting pressure (absolute)

5 bar