Loadstar Sensors DS-3000 Digital Load Cell Display & Controller User Manual

Page 25

Advertising
background image

DS-3000U

DS3000 User Guide

Loadstar Sensors, Inc.

Page: 25 / 48

033-01419 Rev. 4

pnt2

pnt2ta Activate output 1 above this point

pnt2tb Active output 1 below this point

pnt2ra Reset output 1 above this point

pnt2rb Reset output 1 below this point

sen1

s

 or “.”

Include Sensor 1 in Sum with other sensors “S”, individually
“I”, or not included “ ” (blank).

sen2

s

or " "

Include Sensor 2 in Sum with other sensors “S”, individually
”I”, or not included “ ” (blank).

sen3

s

or " "

Include Sensor 3 in Sum with other sensors “S”, individually
“I”, or not included “ ” (blank).

sen4

s

or " "

Include Sensor 4 in Sum with other sensors “S”, individually
“I”, or not included “ ” (blank).

duratn 0.00,

contIn

Active trigger Duration interval (seconds), or continuous.
Setting the duration to 0 causes the DS-3000 to be in
continuous mode, output is active whenever the threshold
conditions are satisfied.

Summary:

Contact output 1 can be used to control external equipment attached by means of
the provided removable wiring connector. Contact terminals are marked [COM --
common], [NO-- normally open] and [NC – normally closed]. The contact terminal
labels indicate the inactive contact state. Ie. [NO] contacts are OPEN when
inactive, but CLOSED when active. Both [NO] or [NC] contacts are referenced to
the [COM] or common terminal.

Outputs may be pulsed by selection of appropriate momentary contact duration,
or continuous. More details and examples programming these entries will be
given in the programming examples section, following later in this document.

8.2.5

Relay Output 2 Menu

Function:

Setup output contact 2 related parameters

Keystrokes:

Press menu till reLay 2 is displayed, then press up or down to select which
parameter you’d like to adjust, then press enter to edit each entry to the desired
value.

These parameters affect Relay Output 2 only! Relay 1 and Relay 2 are
independent, and are programmed separately.

trgr 1

First trigger action threshold. Select enter to enter the
numeric entry field. up and down changes the value of the
currently highlighted digit, and menu changes the currently
highlighted digit. Press enter when finished

Pnt1

pnt1ta Activate output 2 above this value, ie. “Trigger Above”

pnt1tB Activate output 2 below this value, ie. “Trigger Below”

pnt1ra Reset output 2 above this value, ie. “Reset Above”

pnt1rb Reset output 2 below this value, ie. “Reset Below”

Trgr 2

Second hysteresis action threshold. . Select enter to enter
the numeric entry field. up and down changes the value of
the currently highlighted digit, and menu changes the
currently highlighted digit. Press enter when finished