3.1 Preparing the ultrafiltration cell

■■■■■■■■■■■■■■■■■■■■■■

8

■■■■■■■■

IC equipment: Inline-Ultrafiltration

3 Disassembling the ultrafiltration cell

Remove the upper part of the ultrafiltration cell.

Remove the sealing ring.

4 Cleaning the ultrafiltration cell

CAUTION

Damage to the ultrafiltration cell

Organic solvents (e.g. acetone) corrode and damage the ultrafiltra-
tion cell material (PMMA).

Use only ultrapure water or a water-ethanol mixture (70:30) for
cleaning the ultrafiltration cell.

Thoroughly rinse off the sealing ring (2-3), the upper chamber
(2-2) and the lower chamber (2-4) of the ultrafiltration cell with
ultrapure water.

Dry all parts with a lint-free cloth.