Verifying Column Parameters

Remember that commands can be upper or lower case, and that they must always be
followed by <enter>.

To verify that the column parameters have been correctly loaded into the FTC, type
SC<enter> (for “show column”). The column parameters should match those on the data
sheet supplied with the column.

: sc

Column id/sn:

12345

Column serial number

Max. temperature:

325

Maximum column temperature in °C

Calibration data:

Temp.

Res.

40.00

6.30000

The three resistance vs. temperature

183.00

11.74200

calibration points in °C and ohms

325.00

19.65400

Element equation:

R = 1.31039E1 + 6.44364E-2 * T + 1.74357E-4 * T^2

Resistance equation

Input scaling:

V = -2.82099E-3 + 6.00694E-6 * C - 3.83986E-16 * C^2

Voltage equation

I = 2.59489E-4 + 6.02032E-7 * C - 6.12730E-16 * C^2

Current equation

Verifying Control

1. Read the column temperature by typing RT<enter> (for “read temperature”). The

column temperature should be appropriate for its environment. If the temperature
response is not reasonable, recheck the column parameters for errors.

2. Check the column resistance by typing RR<enter> (for “read resistance”).

Getting Started