Detector electrical connections, Initial power-up – VICI D-4-I-SH14-R User Manual

Page 14

Advertising
background image

10

Detector Electrical Connections

CAUTION: Do not use a wrench to tighten the SMC
connectors on the bias and electrometer cables.
Connections should be finger tight only.

1. Referring to Figures 5 and 6 as necessary, connect the BIAS cable to

the top electrode and the electrometer cable (ELECT) to the bottom
electrode.

2. Connect the high-voltage cable from the detector to the pulse supply.

Initial Power-Up

CAUTION: Always make sure that discharge gas is
flowing before heating and powering up the detector.

1. Before installing the column, set the gas flow to 30 ml/min (measured at

the detector vent). Let it flow for 15 minutes so that all air is purged from
the helium purifier.

2. Plug in the helium purifier and turn on the GC.

PULSER

MODULE

ELECTROMETER CABLE

P/N: I-23230-SH14

BIAS CABLE

P/N: I-23404-SH14

SHIMADZU

ELECTROMETER

BOARD

TO 110/220 VAC

PULSER

POWER

SUPPLY

Figure 6: Schematic of electrical connections