System requirements, Components not included with the detector system, System purity – VICI TCD2 User Manual

Page 11: Recommended carrier gas purifiers, Carrier gas selection

Advertising
background image

7

System Requirements

Components Not Included with the Detector System

Carrier gas (99.999% purity is recommended)

Ultra high purity grade gas pressure regulator with stainless steel
diaphragm (recommended)

Any special adapters required for connection to the gas regulator

Flow measuring device

Flow regulating device

System Purity

Since detection of low concentrations depends in part on the purity of the
carrier, the purest carrier available must be used in order to achieve the
lowest possible detection limit. To maintain carrier purity, extra care must be
taken to assure that the delivery system is clean and free of leaks. Use
stainless steel tubing (cleaned to removed manufacturing solvents) instead

of nylon or PTFE tubing, which can diffuse contaminants into the carrier.

Recommended Carrier Gas Purifiers

The Valco Helium Purifer (product number HP2) and Nitrogen Purifer (product
number NP2), which utilize a rare earth gettering alloy to effectively remove
contaminants, are recommended for ppm level analysis of permanent gases.
For other applications, economical and convenient VICI Mat/Sen purifiers are
appropriate. Order the P300-1 for nitrogen, P200-1 for hydrogen, or the
P100-1 for helium and other inert carrier gases.

Carrier Gas Selection

The detector’s response to a component is based upon the difference
between the thermal conductivities of the component and the carrier gas: the
greater the difference, the greater the response. The table below shows
thermal conductivities for a variety of light gases.

Hydrogen 45.9
Helium 36.9
Neon 11.8
Methane 8.6
Oxygen 6.6
Air 6.4
Nitrogen 6.4
Carbon monoxide 6.2
Water 4.5
Argon 4.5
Carbon dioxide 4.2

As an example, note in the table that the thermal conductivity of hydrogen is
46 and helium is 37, while nitrogen is only 6 and argon is even lower at 4.
Since the largest difference in thermal conductivity yields the best response,
detection of small amounts of hydrogen is better done with argon or nitrogen
carrier than with helium.