Dillon Quick-Check Wire Tension Meter (Red) User Manual

Page 20

Advertising
background image

20

Quick-Check Red User Instructions

Comm Port:

This allows you to choose where to export the "ON PRINT" logged
data.

a. Cell = Setup Cell for data export (4 pin Lemo)

b. Com1 = Not used

c. RADIO = Future development

d. Disabled = to turn off

Timed: This is used when the operator wants to record a live load/pull and
store it internally. Can store up to 255 force readings.

Log Rate:

The speed at which the logging of force will be recorded. Based
on seconds (Example 5 sec = Unit will record the force once every
5 seconds)

Log Count:

This is how many logs the operator wants to record. Range is 0 to
255 recorded readings.

Auto Send: Enable/Disable

Can export the readings live if enabled.

Comm Port:

This allows you to choose where to export the "ON PRINT" logged
data.

a. Cell = Setup Cell for data export (4 pin Lemo)

b. Com1 = Not used

c. RADIO = Future development

d. Disabled = to turn off

Overload: This is used when an operator or supervisor wants to know if
their equipment is being overloaded. This will be a percentage based on
capacity and will be below 100% of the Quick Check. (This not the same
as the 120% overload that the Quick Check records based on capacity.)
Maximum Records = 255

Threshold %: (based on capacity)

This the percentage above where the unit will record the peak
load.

Re-arm %: (based on capacity)

This is the load percentage the force must drop below before the
"OVERLOAD" will reset (re-arm) and store another load.

Holdoff Seconds:

Once the unit has recorded this overload, how much time will pass
before this unit will re-arm and allow to record the overload again.