DT 200 Users Manual version 1.E

Diagnostic Tests

30

Access Parameter Functions

Hold the CLEAR key down and press the ENTER key. The numeric display prompts

“CFG xx” where xx is the currently selected parameter. Enter a parameter function
number and then press the ENTER key. NOTE: use the UNITS key (

↓) and the PRINT

key (

↑) to scroll the CFG parameter number. Press the UNITS key to increment the CFG

number, press the PRINT key to decrement the CFG number.
To exit configuration and store changes, press the CLEAR key when “CFG xx” is being

displayed.
To exit configuration and NOT store changes, enter 999 when “CFG xx” is being

displayed. The display prompts “Abort”. Press the ENTER key to abort (exit without

saving changes) or press the CLEAR key if you do not want to abort.

Example: Begin Calibration of the Scale-
The DT200 is in the display weight mode. Hold the CLEAR key down and press the

ENTER key to enter the configure mode. The DT200 prompts “CFG xx” where xx is

the last selected parameter. Enter 60 and then press the ENTER key. The DT200

prompts “dEAdld”. Begin the calibration process or press the CLEAR key to exit.

Function 59 Display Calibration Audit Number.

Use function 59 to view the audit and configuration numbers, the software ID number,

and the status of the internal function lock.
The DT200 provides a calibration audit number and a configuration number. They are

used to indicate changes to calibration data and configuration data and can be used in

place of mechanical seals by some regulatory agencies. The DT200 audit number

increments each time a calibration parameter is changed (deadload or weight conversion

factor). The configuration number increments each time a password level 3 parameter is

changed. If physical seals are used, then place the internal function lock in the locked

position before sealing. Calibration and password level 3 parameters cannot be accessed

when the function lock is in the locked position.
1. Hold the CLEAR key down and press the ENTER key to enter the configure mode.

2. Enter CFG 59. The DT200 prompts “A xxxx” where xxxx is the audit number.

3. Press the ENTER key. The DT200 prompts “C xxxx” where xxxx is the current

calibration number.

4. Press the ENTER key. The DT200 prompts “bixxxx” where bixxxx is the ID number

of the software installed in the DT200.

5. Press the ENTER key. The DT200 prompt “un-LoC” if the function lock is in the

unlocked position, or “Loc Ed” if the function is in the locked position.