HCGG Maintenance Checklist

Item

Process

Start System

First, start the combine. Next, turn on HarvestMaster System Console (Handheld should turn on

automatically). Finally, load FRS.

Air Pressure

Air pressure should be set 60 –80 psi.

Air Leaks

If you have a reservoir air tank, turn off the combine after filling the tank. Check for air leaks on all

cylinders and solenoids. Also check solenoid exhaust port plugs for any build-up of oil or debris. If

they are caked in debris, take the exhaust port out and clean with Brake & Parts Cleaner.

Cable Connections

Ensure all cable connections are tight and there is not any debris in the connections.
Be careful when reconnecting the cables not to bend or damage any connector pins.

Debris

Blow out the Graingage at the end of every day. DO NOT use high pressure washer to clean system.

Bucket Clearance

The plot bucket should be sitting firmly on the load cell tracks. Check to ensure that all cables and

air hoses are not touching the bucket. Also, check the clearance around the bucket.

Actuators Cycle Smoothly

Make sure all cylinders open and close smoothly.

Weight System

Place a known weight into the weight bucket. Verify that the system reads the weight accurately.

Remove the weight and verify the system zeros out.

Calibrate Bulk Density Sensor

(BDS)

Procedure is in the manual (HCGG pg. 67).

Calibrate Test Weight

Procedure is in the manual (HCGG pg. 48).

Calibrate Slope and Motion

Procedure is in the manual (HCGG pg. 29).

Calibrate Moisture Curve

Procedure is in the manual (HCGG pg. 38).

HCGG System Calibrating Guide

Load Cells—Calibration

1. Select Setup > Hardware Setup > HCGG Setup > Weight Calibration

Slope and Motion Sensor needs to be turned OFF during calibration.

2. Follow instructions on handheld screen.

BDS Test Weight Calibration

1. Select Setup (F3) > Hardware Setup > HCGG Setup > Test Weight.
2. Enter diagnostics screen by selecting Diag (F3).
3. Tare the BDS load cell by selecting Tare LC (F2).
4. Place BDS weight in cup.
5. Use the formula below to figure out how much to adjust the Load Cell Multiplier.

New Load Cell Multiplier =Actual Weight/BDS Measured Weight * Load Cell Multiplier

6. Enter new Load Cell Multiplier and save.

BDS Cup Volume Adjustment

1. Select Setup (F3) > Hardware Setup > HCGG Setup > Test Weight.
2. Enter diagnostics screen by selecting Diag (F3).
3. Tare the BDS load cell by selecting Tare LC (F2).
4. Pour a grain sample with a known test weight into the weigh bucket.
5. Press Level (F3) to level the grain to the top of the cup. Observe the Test Weight reading in the diagnostics.
6. If the test weight reading of the BDS does not match the known test weight of the grain sample, adjust cup volume using the formula

below.

New Cup Volume = BDS Measured TW / Actual TW from Standard * Cup Volume