Immunity test

Voltage dips, short

interruptions, and

voltage variations

on power supply

input lines

IEC 61000-4-11

Power frequency

(50/60 Hz) magnetic

field

IEC 61000-4-8

Conducted RF

IEC 61000-4-6
Radiated RF

IEC 61000-4-3

IEC 60601 test
level

< 5% U

T

(>95% dip in U

T

)

for 0.5 cycle
40% U

T

(60% dip in U

T

)

for 5 cycles
70% U

T

(30% dip in U

T

)

for 25 cycles
<5% U

T

(>95% dip in U

T

)

for 5 sec
3 A/m

3 Vrms 150 kHz

to 80 Mhz
20 V/m 80 Mhz

to 2.5 Ghz

Electromagnetic
environment/guidance

Mains power quality should

be that of a typical

commercial or hospital

environment. If the user

of the Portable Oxygen

Concentrator required

continued operation during

power main interruptions,

it is recommended that the

concentrator be powered

from an uninterruptible

power supply or battery.

Power frequency magnetic

fields should be at levels

characteristic of a typical

commercial or hospital

environment.
Portable and mobile RF

communications equipment

should be used no closer

to any part of the Portable

Oxygen Concentrator,

including cables, than the

recommended separation

distance calculated from the

equation applicable to the

frequency of the

transmitter.
Recommended separation

distance:
d = 1.2

√P

d = 1.2

√P 80 MHz to 800

MHz
d = 2.3

√P 800 MHz to 2.5

GHz
where P is the maximum

output power rating of the

transmitter in watts (W)

according to the transmitter

manufacturer and d is the

recommended separation

distance in meters (m).
Field strengths from fixed

RF transmitters, as deter-

mined by an electromagnetic

site survey

a

should be less

than the compliance level

in each frequency range

b

.

Interference may occur in the

vicinity of equipment marked

with the following symbol:

Compliance
level

< 5% U

T

(>95% dip in U

T

)

for 0.5 cycle
40% U

T

(60% dip in U

T

)

for 5 cycles
70% U

T

(30% dip in U

T

)

for 25 cycles
<5% U

T

(>95% dip in U

T

)

for 5 sec
3 A/m

3 Vrms

20 V/m

Note: At 80 MHz and 800 MHz, the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures,
objects, and people.

a

Field strengths from fixed transmitters, such as base stations for radio

(cellular/cordless) telephones and land mobile radios, amateur radio, AM
and FM radio broadcast, and TV broadcast cannot be predicted theoretical-
ly with accuracy. To assess the electromagnetic environment due to fixed
RF transmitters, an electromagnetic site survey should be considered. If
the measured field strength in the location in which the Portable Oxygen
Concentrator is used exceeds the applicable RF compliance level above,
the concentrator should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as
reorienting or relocating the concentrator.

b

Over the frequency range 150 kHz to 80 MHz, field strengths should be

less than 3 V/m.

DL-00096 rev 7.indd 24