Replacing the diaphragm – Labconco VACUUBRAND Diaphragm Pump 7393000 User Manual

Page 58

Advertising
background image

page 58 of 74

Opening the hose clip:

Apply slotted screwdriver as shown and

turn.

Detach the coupling of the connection tube (V)

to the other side of the pump at the valve head

(O).

Pump at pump support:

Detach the coupling of the connection tube (V)

to the other side of the pump as well as the

hose connection to the inlet/outlet of the vacu-

um system at the valve head (O).


Open the hose clip (U) with a slotted screw-

driver.

Pull the tubing off the hose connector.

Disassemble head covers (J) to check the

diaphragm (F).

Unscrew four (pump with one/two heads) or

eight (pump with four heads) Allen screws

(H) with a 5mm wide Allen key. Remove both

head covers (J) (pumps with one/two heads:

only one head cover) together with valve

heads (O) and connections.

+

It is not necessary to disassemble the valve

heads (O), the connection fasteners (Q), or

the hose connection between the adjacent

head covers (pumps with four heads).

Replacing the diaphragm

V

O

U

J

H

O

Advertising
This manual is related to the following products: