1 general description – Labconco Guardian 500 Airflow Monitor User Manual

Page 4

Advertising
background image

4

1.1 General

Description

All systems comprise of the following components :-

1 – Guardian 500 Alarm unit,
1 – AC power supply
1 – Inner sidewall adaptor

If the Sash Alarm System option is included there will also be a sash micro switch or proximity
switch.

Operator Features --- the alarm has the following operator features :-



Normal Airflow - Green LED (Not flashing) will be displayed if the airflow is
greater than the Low air alarm point.

Low Airflow - Red LED (flashing) will be displayed if the airflow is lower
than the Low air alarm point.


Sash High - Amber LED (Flashing) will be displayed when the Sash is raised
above the max safe working opening.

Audible Alarm -- the Audible alarm will sound ( can be muted ) in the Air Low
and the Sash High alarm condition

Night Set-back -- when the Night Setback input is activated the Audible alarm
will be muted and the Setback LED will illuminate.

External Alarm -- when the External Alarm input is activated the Audible alarm
will sound and the External Alarm LED will illuminate.



















Advertising